Abstract

© Copyright 2016, American Society for Microbiology. All Rights Reserved. Following revised information pertaining to newer swab types and testing protocols in the new CLSI M40-A2 standard, we evaluated three liquid swab transport systems for the recovery of aerobic, anaerobic, and fastidious organisms at room temperature and at 4°C. All tested liquid swab transport systems were fully compliant with the M40-A2 standard, with acceptable performance at both temperatures after the full specified holding period, using both qualitative (roll-plate) and quantitative (swab elution) methods.

Citation

Gizzie, N., & Adukwu, E. (2016). Evaluation of liquid-based swab transport systems against the new approved CLSI M40-A2 standard. Journal of Clinical Microbiology, 54(4), 1152-1156. https://doi.org/10.1128/JCM.03337-15

Journal Article Type Article
Acceptance Date Jan 25, 2016
Online Publication Date Feb 3, 2016
Publication Date Apr 1, 2016
Deposit Date May 14, 2018
Publicly Available Date May 14, 2018
Journal Journal of Clinical Microbiology
Print ISSN 0095-1137
Electronic ISSN 1098-660X
Publisher American Society for Microbiology
Peer Reviewed Peer Reviewed
Volume 54
Issue 4
Pages 1152-1156
DOI https://doi.org/10.1128/JCM.03337-15
Public URL https://uwe-repository.worktribe.com/output/912590
Publisher URL http://dx.doi.org/10.1128/JCM.03337-15

Files





Related Outputs



You might also like



Downloadable Citations