Skip to main content

Research Repository

Advanced Search

Application of screen-printed microband biosensors to end-point measurements of glucose and cell numbers in HepG2 cell culture

Jackson, S. K.; Pemberton, Roy; Xu, Jinsheng; Hart, John P.; Pittson, R.; Biddle, N.; Drago, G. A.

Authors

S. K. Jackson

Jinsheng Xu

R. Pittson

N. Biddle

G. A. Drago



Abstract

Microband glucose biosensors were produced by insulating and sectioning through a screen-printed, water-based carbon electrode containing cobalt phthalocyanine redox mediator and glucose oxidase enzyme. Under quiescent conditions at 37 °C, at an operating potential of +0.4 V, they produced an amperometric response to glucose in buffer solutions with a sensitivity of 26.4 nA/mM and a linear range of 0.45 to 9.0 mM. An optimal pH value of 8.5 was obtained under these conditions, and a value for activation energy of 40.55 kJ mol-1 was calculated. In culture medium (pH 7.3), a sensitivity of 13 nA/mM was obtained and the response was linear up to 5 mM with a detection limit of 0.5 mM. The working concentration was up to 20 mM glucose with a precision of 11.3% for replicate biosensors (n = 4). The microband biosensors were applied to determine end-point glucose concentrations in culture medium by monitoring steady-state current responses 400 s after transfer of the biosensors into different sample solutions. In conjunction with cultures of HepG2 (human Caucasian hepatocyte carcinoma) cells, current responses obtained in 24-h supernatants showed an inverse correlation (R2 = 0.98) with cell number, indicating that the biosensors were applicable for monitoring glucose metabolism by cells and of quantifying cell number. Glucose concentrations determined using the biosensor assay were in good agreement, for concentrations up to 20 mM, with those determined spectrophotometrically (R2 = 0.99). This method of end-point glucose determination was used to provide an estimated rate of glucose uptake for HepG2 cells of 7.9 nmol/(106 cells min) based on a 24-h period in culture. © 2008 Elsevier Inc. All rights reserved.

Citation

Jackson, S. K., Pemberton, R., Xu, J., Pittson, R., Biddle, N., Drago, G. A., & Hart, J. P. (2009). Application of screen-printed microband biosensors to end-point measurements of glucose and cell numbers in HepG2 cell culture. Analytical Biochemistry, 385(2), 334-341. https://doi.org/10.1016/j.ab.2008.10.037

Journal Article Type Article
Publication Date Feb 15, 2009
Journal Analytical Biochemistry
Print ISSN 0003-2697
Electronic ISSN 1096-0309
Publisher Elsevier Masson
Peer Reviewed Peer Reviewed
Volume 385
Issue 2
Pages 334-341
DOI https://doi.org/10.1016/j.ab.2008.10.037
Keywords glucose biosensor, microband electrode, cell culture, HepG2 cells, glucose metabolism
Public URL https://uwe-repository.worktribe.com/output/1005005
Publisher URL http://dx.doi.org/10.1016/j.ab.2008.10.037