A new approach to the determination of column overload characteristics in reversed-phase liquid chromatography
Dai, Jun; Carr, Peter W.; McCalley, David V.
Authors
Abstract
Column overloading is very common during the separations of basic analytes in analytical scale reversed-phase liquid chromatography (RPLC). Due to the complex interactions of ionic analytes with stationary and mobile phases, only a very small amount of ionized sample compared to the amount of nonpolar solute can be injected before the peak shape is distorted by non-linear chromatographic processes. Often the amount that can be injected before overload is observed is so small that the signal is quite noisy, thereby making the measured plate count imprecise and possibly inaccurate. The purpose of the present study was to develop a practical method for the precise measurement of the plate count and a column overload parameter using a simple but mathematically rigorous model of Langmuirian non-linear chromatography. An "overload profile", i.e. a plot of apparent plate count versus amount injected, is characterized by two parameters: the limiting plate count (N 0 ) and the column sample loading capacity (ω 0.5 ). The limiting plate count is the plate count that should be observed when the amount of sample injected is so small that a linear isotherm pertains. The column sample loading capacity, which is taken as the sample load that leads to a plate count equal to half of the limiting plate count, is a measure of the maximum amount of sample that can be injected into that column. The approach was tested by applying it to the study of cationic analytes in RPLC. We show that N 0 under constant conditions (column length, flow rate, mobile phase composition, etc.) is almost independent of column type (manufacturer); however, different column types (at the same length, diameter and flow rate) exhibit clear differences in their sample loading capacity (ω 0.5 ). We believe that for most well packed type B columns, the column sample loading capacity and not the limiting plate count is the more important property that accounts for most of the apparent differences in peak width when different types of columns are examined. © 2009 Elsevier B.V. All rights reserved.
Journal Article Type | Article |
---|---|
Publication Date | Mar 20, 2009 |
Journal | Journal of Chromatography A |
Print ISSN | 0021-9673 |
Publisher | Elsevier |
Peer Reviewed | Peer Reviewed |
Volume | 1216 |
Issue | 12 |
Pages | 2474-2482 |
DOI | https://doi.org/10.1016/j.chroma.2009.01.036 |
Keywords | high performance liquid chromatography, overload langmuirian non-linear chromatography, limiting plate count, column sample loading capacity |
Public URL | https://uwe-repository.worktribe.com/output/997794 |
Publisher URL | http://dx.doi.org/10.1016/j.chroma.2009.01.036 |
You might also like
Managing sample introduction problems in hydrophilic interaction liquid chromatography
(2023)
Journal Article
Downloadable Citations
About UWE Bristol Research Repository
Administrator e-mail: repository@uwe.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search