An electrochemical sensor device for measuring blood ammonia at the point of care
Brannelly, N. T.; Killard, Anthony
Abstract
The level of ammonia in blood is relevant in a number of medical conditions. While ammonia is a marker of dysfunction, elevated ammonia is itself a serious medical emergency and can lead to significant and permanent neurological impairment if not addressed quickly. Blood ammonia testing is typically performed in the central laboratory. While a number of point of care devices have been developed, these are based on classical enzymatic or colorimetric principles and have not been widely adopted. In this work, an electrochemical sensor device was developed for measuring blood ammonia. The device was based on the deposition of polyaniline nanoparticle films onto screen printed interdigitated electrodes using inkjet printing and their integration into a polymer microfabricated device with a polytetrafluoroethylene membrane. The device required a 52 µL serum sample and measured the change in impedance of the sensor with respect to air at 1 kHz, 5 mV rms. The device was capable of the measurement of ammonia in serum across the physiologically relevant range of 25–200 µM (r2 =0.9984) and had a limit of detection of 12 µM (n =3). The device showed no significant issues with common electrochemical interferences in blood. The device was also validated against a commercial spectrophotometric assay which resulted in excellent correlation (r =0.9699, p < 0.0001) between both methods (n =3). When stored under desiccation, devices displayed minimal variation over time (0.64%) with respect to their impedance in air (n =12) and could be stored in desiccant for at least five months.
Journal Article Type | Article |
---|---|
Acceptance Date | Feb 10, 2017 |
Online Publication Date | Feb 13, 2017 |
Publication Date | May 15, 2017 |
Deposit Date | Feb 14, 2017 |
Publicly Available Date | Feb 13, 2018 |
Journal | Talanta |
Print ISSN | 0039-9140 |
Publisher | Elsevier |
Peer Reviewed | Peer Reviewed |
Volume | 167 |
Pages | 296-301 |
DOI | https://doi.org/10.1016/j.talanta.2017.02.025 |
Keywords | Ammonia, blood, electrochemical, sensor, point-of-care |
Public URL | https://uwe-repository.worktribe.com/output/887711 |
Publisher URL | http://dx.doi.org/10.1016/j.talanta.2017.02.025 |
Contract Date | Feb 14, 2017 |
Files
1-s2.0-S0039914017302394-main.pdf
(841 Kb)
PDF
Blood ammonia sensor_Talanta_REVISION.docx
(191 Kb)
Document
You might also like
Disposable sensors
(2017)
Journal Article
Downloadable Citations
About UWE Bristol Research Repository
Administrator e-mail: repository@uwe.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search