Abstract
Purpose
The aim of the study was to determine the test-retest reliability of lower limb kinematic waveforms derived from 3D gait analysis (3DGA) in patients following total hip arthroplasty (THA).
Methods
Eight (7M:1F; age: 70 ± 7 years; height: 1.68 ± 0.11m; mass: 85 ± 20kg) adults with a unilateral THA attended test and retest sessions. 3DGA was undertaken with participants walking at a self-selected pace along a 7m walkway within each session. The standard error or the measurement (SEM) was calculated for hip, knee and ankle
joint angles in all three planes, over the walking gait cycle.
Results
The SEM ranged from 2.9 – 4.1°, 2.7 – 3.7° and 1.9 – 3.9°, in the sagittal, frontal and traverse planes at the hip. At the knee the SEM ranged from 1.6 – 4.2°, 1.0 – 1.9° and 1.3 – 2.9° in the sagittal, frontal and transverse planes, respectively. While the SEM
ranged from 0.7 – 2.0°, 1.2 – 2.3° and 2.9 – 4.0° in the sagittal, frontal and transverse planes at the ankle.
Conclusions
The findings demonstrate that 3DGA provides a reliable means of quantifying lower limb kinematics over the walking gait cycle in patients following THA, with all SEM values below the 5° threshold previously suggested to identify clinically meaningful differences. The SEM values reported may aid in the interpretation of changes in lower limb kinematics in patients following THA.
Citation
Langley, B., Greaves, H., Whelton, C., Page, R., Cramp, M., Morrison, S. C., …Board, T. (2024). Reliability of kinematic waveforms during gait analysis with total hip arthroplasty patients. Journal of Joint Surgery and Research, 2, 1-6. https://doi.org/10.1016/j.jjoisr.2023.11.003
Journal Article Type | Article |
---|---|
Acceptance Date | Nov 23, 2023 |
Online Publication Date | Dec 6, 2023 |
Publication Date | Mar 31, 2024 |
Deposit Date | Jan 18, 2024 |
Publicly Available Date | Jan 18, 2024 |
Journal | Journal of Joint Surgery and Research |
Publisher | Elsevier |
Peer Reviewed | Peer Reviewed |
Volume | 2 |
Pages | 1-6 |
DOI | https://doi.org/10.1016/j.jjoisr.2023.11.003 |
Public URL | https://uwe-repository.worktribe.com/output/11618559 |
Additional Information | This article is maintained by: Elsevier; Article Title: Reliability of kinematic waveforms during gait analysis with total hip arthroplasty patients; Journal Title: Journal of Joint Surgery and Research; CrossRef DOI link to publisher maintained version: https://doi.org/10.1016/j.jjoisr.2023.11.003; Content Type: article; Copyright: © 2023 The Author(s). Published by Elsevier B.V. on behalf of The Japanese Hip Society, The Japanese Society for Replacement Arthroplasty and The Japanese Knee Society. |
Files
Reliability of kinematic waveforms during gait analysis with total hip arthroplasty patients
(2.9 Mb)
PDF
Licence
http://creativecommons.org/licenses/by-nc-nd/4.0/
Publisher Licence URL
http://creativecommons.org/licenses/by-nc-nd/4.0/
You might also like
Downloadable Citations
About UWE Bristol Research Repository
Administrator e-mail: repository@uwe.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search