Skip to main content

Research Repository

Advanced Search

Practical assessment of frictional heating effects and thermostat design on the performance of conventional (3 μm and 5 μm) columns in reversed-phase high-performance liquid chromatography

Fallas, Morgane M.; Hadley, Mark R.; McCalley, David V.

Authors

Morgane M. Fallas

Mark R. Hadley



Abstract

A practical investigation of frictional heating effects in conventional C18 columns was undertaken, to investigate whether problems found for sub-2 μm columns were also present for those of particle size 3 μm and 5 μm and different internal diameter. The influence of a water bath, a still air heater, and a forced air heater on performance was investigated. Heating effects were substantial, with a decrease in k of almost 15% for toluene over the flow rate range ∼0.4-2.3 mL/min with a 15 cm × 0.46 cm ID column packed with 3 μm particles. Heating effects on retention increased with increasing solute k, with increase in the column ID, with decrease in the column particle size, and with decrease in the set column oven temperature. While the water bath minimised axial temperature gradients and thus its effect on k, radial temperature gradients were potentially serious with this system, especially at high mobile phase velocity, even with columns containing 5 μm particles. In contrast to the effects of axial temperature gradients in 4.6 mm columns, very little difference in Van Deemter plots was noted between the three different thermostats with 2 mm ID columns, even when 3 μm particles were used. However, the efficiency of 2 mm columns for peaks of low or moderate k (k < 4) can be compromised by the extra dead volume introduced by the heating systems, even with conventional HPLC systems with otherwise minimised extra column volume. © 2009 Elsevier B.V. All rights reserved.

Citation

Fallas, M. M., Hadley, M. R., & McCalley, D. V. (2009). Practical assessment of frictional heating effects and thermostat design on the performance of conventional (3 μm and 5 μm) columns in reversed-phase high-performance liquid chromatography. Journal of Chromatography A, 1216(18), 3961-3969. https://doi.org/10.1016/j.chroma.2009.03.007

Journal Article Type Article
Publication Date May 1, 2009
Journal Journal of Chromatography A
Print ISSN 0021-9673
Publisher Elsevier
Peer Reviewed Peer Reviewed
Volume 1216
Issue 18
Pages 3961-3969
DOI https://doi.org/10.1016/j.chroma.2009.03.007
Keywords HPLC, frictional heating, retention, column efficiency
Public URL https://uwe-repository.worktribe.com/output/996219
Publisher URL http://dx.doi.org/10.1016/j.chroma.2009.03.007